Electrodermal Activity
Much of this information was gleaned from the following sources:
- sdfasd
- sdfasdf
- asdfasd
There are several aspects of the EDA that one can measure:
- SC = Skin conductance. Measured in microsiemens or micromhos (the reciprocal of resistance, ohm)
- SCL = tonic changes in SC
- note: we will generally not focus on SCL for a number of reasons. These include the fact that SCL will drift over time due to both physiological (SCRs can raise SCL) and non-physiological (polarization of electrodes in DC recording) factors.
- SCR = phasic changes in SC
- NS-SCR = SCRs that occur without any identifiable eliciting stimulus
- ER-SCR = SCR that is evoked by a specific eliciting stimulus
The onset of an ER-SCR will occur 1 to 3 seconds after the event and will have a rise-time (i.e. from initial deflection to peak amplitude) of 0.5 to 5 seconds.
10% of participants will be non-responders. That is, these individuals will not show an appreciable SCR.
Moreover, there are several ways to evaluate these responses. With regard to SCRs, one can
- evaluate the peak amplitude of the SCR
- evaluate the rise-time of the SCR
- evaluate the recovery time of the SCR
- note: generally this is measured to 67% or 50% recovery because after an SCR there will likely be an increase in SCL. That is, the response will not completely recover to pre-SCR baseline.
- evaluate the frequency of NS-SCRs within a given epoch
- evaluate the rate of SCR habituation
Recording Electrodermal Responses
Useful links
Initial Setup
Electrode Setup
- Connect two LEAD110A leads into the EDA100C
- Plug one lead into the
Vin+
port and the other into theVin-
port
- Use electrode prep swab to clean the index and middle fingers of the participant's non-dominant hand
- Remove two EL507 electrodes.
- Put extra GEL101 on them
- You must use the specific SCR GEL101. Other conductive gels (e.g. the EEG gel) are not appropriate.
- Attach the pinch connector of one LEAD110A to the EL507 electrode on the middle finger, and the other to the EL507 electrode on the pointer finger.
- Wait at least five minutes before recording. The electrolyte gel will continue to penetrate into the stratum corneum, which will cause a drift in the ECL (specifically, increased conductivity).
kpnl/biopac/scr.1445952633.txt.gz · Last modified: 2015/10/27 08:30 by admin